Comparison of Fixation Stability of Resorbable Versus Titanium Plate and Screws in Mandibular Angle Fractures

Burak Bayram, DDS, PhDCorresponding Author Informationemail address, Kenan Araz, DDS, PhD, Sina Uckan, DDS, PhD, Cenk Balcik, PhD§

Purpose

The aim of this study was to compare the fixation reliability and stability of titanium and resorbable plates and screws by simulating chewing forces.

Materials and Methods

Mandibular angle fractures in 11 sheep hemimandibles were fixed with 4-hole straight titanium plates and 2.0 × 7–mm titanium screws; in addition, 11 hemimandibles were fixed with 4-hole straight resorbable plates and 2.5 × 6–mm resorbable screws according to the Champy technique. The hemimandibles were mounted with a fixation device in a servohydraulic testing unit for compressive testing. Displacement values under 20, 60, 100, 120, 150, and 200 N; maximum displacements; and maximum forces that the model could resist before breakage were recorded and compared.

Results

Significant differences were found between resorbable and titanium plates and screws at all forces (20, 60, 100, 120, 150, and 200 N) (P < .05). We found no statistically significant differences in the breaking force and maximum displacement values (displacement values at the breaking forces) between the groups.

Conclusions

The stability of mandibular angle fractures with titanium miniplates under simulated chewing forces was significantly higher than with the resorbable system. Metallic and resorbable fixation systems cannot be used interchangeably to treat mandibular angle fractures under similar loading conditions.

 Fellow, Department of Oral and Maxillofacial Surgery, School of Dentistry, Baskent University, Ankara, Turkey

 Professor, Department of Oral and Maxillofacial Surgery, School of Dentistry, Baskent University, Ankara, Turkey

 Professor, Department of Oral and Maxillofacial Surgery, School of Dentistry, Baskent University, Ankara, Turkey

§ Assistant Professor, Department of Mechanical Engineering, Baskent University, Ankara, Turkey

Corresponding Author InformationAddress correspondence and reprint requests to Dr Bayram: Baskent Universites Dishekimligi Fakultesi, 11.sok No: 26 06490 Bahcelievler, Ankara, Turkey

 This study was supported and funded by the Baskent University Academic Research Projects Unit.

PII: S0278-2391(09)00301-2

doi:10.1016/j.joms.2009.03.011


View previous. 16 of 51 View next.